    Description

    High Vacuum Bakeout Chambers
    Custom Built Dual Piece Part Vacuum Bake System used to de-gas optical sensor components at a high temperature (up to
    450
    o
    C) and high vacuum (5.0 E-7 Torr or better).
    Major Components include: ISP-500B Anest Iwata oil-free scroll vacuum pump, CTI On-Board 8 Cryopump with CTI 8200 water cooled 3 phase Compressor, MKS Vacuum Gauges, Watlow Temperature, Fail-Safe & Power Controllers. Allen-Bradley PLC & Control Equipment includes IDEC Slim Line Power Supplies, 5-Port Ethernet Switch & touch-screen computer. The casters have leveling feet.
    Each Chamber is Constructed of 304L Electro-Polished Stainless Steel.
    Each Chamber is 12" Diameter x 18" Deep & is vertically mounted with a full opening hinged lid.
    Each lid is flat, includes a Viton O-ring inside a water-cooled O-ring groove with a latched handle
    and mechanical stop to prevent the lid from opening more than 30
    o
    beyond vertical.
    Electro-Polished Stainless Steel adjustable parts tray assembly is included to be utilized for parts placement within the chamber.
    Operator Interface with Touch Screen
    Records & Charts Vacuum Pressure & Temperature & displays Vacuum Pressure, Valve Status, Oven Temperature, Heater Status & Current Cryopump Temperature in Degrees, Kelvin.
    Features CTI On-Board 8 Cryopump, Allen Bradley PLC with 3 modes of operation.
    1) Hand Mode: No interlocks
    2) Manual Mode: with interlocks
    3) Automatic Mode: Auto Auto-pump down and Vent.
    Watlow Heater Power Controllers
    PLC Hand Switches
    Allen-Bradley PLC & Electrical Components built by UL Approved Panel Shop
    CTI 8200 Water-Cooled, 3 phase Compressors on elevated frames with casters.
    Facility Requirements:  208VAC / 3 Phase / 60 Hz / 62 FLA, Water Connection of 2 GPM.
    Shop Air or GN2 @ 60 to 80 psi used for valve operation, GN2 @ 10 psi used for oven vent/backfill.
    Cryopump Compressor Facility Requirements: 208/240 VAC, 20 Amp/3 Phase, Chilled Water 60 to 80 Degrees F @ 2 GPM